pubmed:issue
3
pubmed:dateCreated
2001-8-22
pubmed:abstractText
This study involves the measurement of photoneutron contamination emitted from a Siemens Primus medical linear accelerator by using BD-PND bubble detectors. Various bubble detectors were arranged around the linac head with the interval of I m and at the same height as the isocenter to measure the dose equivalent distribution in the treatment room. The measurements were performed for 15 MV X-rays with 40 x 40cm2 and 0 x 0cm2 fields and for 15,18, and 21 MeV electrons with 25 x 25 cm2 electron cone. Neutron dose equivalent rate at the points of measurement in the treatment room decreased with increasing distance to the isocenter. The maximum neutron dose equivalents were at the isocenter, and the values for 15MV 40 x 40 and 0 x 0 cm2 were 1843+/-90 and 169.9+/-59.9 microSv per Gy X-ray, respectively. The values for 15, 18 and 21 MeV electrons with 25 x 25 cm2 cones were 100.0+/-20.4, 262.7+/-61.2 and 349.0+/-29.6 microSv per Gy electron, respectively. The neutron contamination of electrons less than 12 MeV was below the detection limit.
pubmed:language
eng
pubmed:journal
pubmed:status
PubMed-not-MEDLINE
pubmed:month
Sep
pubmed:issn
0969-8043
pubmed:author
pubmed:issnType
Print
pubmed:volume
55
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
315-21
pubmed:dateRevised
2003-10-31
pubmed:year
2001
pubmed:articleTitle
The measurement of photoneutrons in the vicinity of a Siemens Primus linear accelerator.
pubmed:affiliation
Department of Nuclear Science, National Tsing Hua University, Hsinchu, Taiwan.