pubmed:abstractText
During cutting of the femoral cavity in the ROBODOC procedure, an aerosol cloud of irrigation fluid, blood, and tissue debris was seen. This cloud potentially is contaminated with bacterial and viral vectors, which is an infection risk for the surgical team. A flat and a ball cutter were tested in four standard situations macroscopically with a colored solution. In a second experiment, the cutter was exposed to a fluid contaminated with Staphylococcus aureus, and bacterial room contamination was detected using standard cultures. The aerosol cloud was seen in a 6 x 3.6-m area. Extension and concentration varied, depending on the irrigation situation. ROBODOC's high-speed cutter produces an aerosol cloud in an area in which all members of the surgical team are affected. Sufficient protection is necessary for everyone in the operating room.