1-2
pubmed:dateCreated
2001-6-6
pubmed:abstractText
An improved high-performance liquid chromatographic method has been developed to measure human plasma concentrations of the analgesic nonsteroidal anti-inflammatory drug ketorolac for use in pharmacokinetic studies. Samples were prepared for analysis by solid-phase extraction using Bond-Elut PH columns, with nearly complete recovery of both ketorolac and the internal standard tolmetin. The two compounds were separated on a Radial-Pak C18 column using a mobile phase consisting of water-acetonitrile-1.0 mol/l dibutylamine phosphate (pH 2.5) (30:20:1) and detected at a UV wavelength of 313 nm. Using only 250 microl of plasma, the standard curve was linear from 0.05 to 10.0 microg/ml.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
1387-2273
pubmed:author
pubmed:issnType
Print
pubmed:day
5
pubmed:volume
755
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
383-6
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Determination of ketorolac in human plasma by reversed-phase high-performance liquid chromatography using solid-phase extraction and ultraviolet detection.
pubmed:affiliation
Department of Anesthesiology, Northwestern University Medical School, Chicago, IL 60611, USA.
pubmed:publicationType
Journal Article, Case Reports, Research Support, Non-U.S. Gov't