pubmed:abstractText
Cloud point extraction has been used for the preconcentration of manganese, after the formation of a complex with 1-(2-thiazolylazo)-2-naphthol (TAN), and later analysis by flame atomic absorption spectrometry using octylphenoxypolyethoxyethanol (Triton X-114) as surfactant. The chemical variables affecting the separation phase and the viscosity affecting the detection process were optimized. Under the optimum conditions (i.e., pH = 9.2, [TAN] = 2.0 x 10(-5) mol l-1, [Triton X-114] = 0.05%, added methanol volume = 0.2 ml), preconcentration of 50 ml of sample solution permitted the detection of 0.28 ppb for manganese. The enhancement factor was 57.6. The proposed method has been applied to the determination of manganese in water samples.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0003-2654
pubmed:author
pubmed:issnType
Print
pubmed:volume
126
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
534-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Determination of manganese in water samples by flame atomic absorption spectrometry after cloud point extraction.
pubmed:affiliation
Division of Chemistry, Natural Sciences, National Institute of Education, Nanyang Technological University, 1 Nanyang Walk, Singapore 637616. kcteo@nie.edu.sg
pubmed:publicationType