pubmed:abstractText
We report our recent experience of using argon plasma to endoscopically cut biliary Wallstent prostheses in these patients. The first patient had a bleeding duodenal ulceration caused by the impaction of the prosthesis meshes whereas the second patient had an ill-positioned biliary stent with impaction into the opposite duodenal wall. Both prostheses were shortened using argon plasma. In the third patient, the lower extremity of a obstructed biliary Wallstent was positioned in the third duodenum preventing its endoscopic catheterization. After shortening using argon plasma, a new plastic stent could be inserted to allow drainage. The outcomes in these cases demonstrate the feasibility of endoscopically shortening metallic Wallstents after release using argon plasma.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
0013-726X
pubmed:author
pubmed:issnType
Print
pubmed:volume
33
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
289-90
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Argon plasma endoscopic section of biliary metallic prostheses.
pubmed:affiliation
Dept. of Gastroenterology, Archet II, University Hospital, Nice, France. demarquay.jf@chu-nice.fr
pubmed:publicationType
Journal Article, Case Reports