pubmed:abstractText
In this study, we evaluated the performance and blood compatibility of polyethersulfone hollow fiber membrane plasma separator by animal experiment. Hemolysis did not occur under the usual conditions of plasma separation. The sieving coefficients of total protein, albumin and globulin were over 95%, and about 60% of total plasma were extracted from the whole blood. White blood cells, platelets, fibrinogen, and coagulation factors were decreased during the early stage of plasma separation and appeared to be within acceptable ranges for clinical use.
pubmed:language
chi
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
1001-5515
pubmed:author
pubmed:issnType
Print
pubmed:volume
17
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
249-54
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
[Evaluation of performance and blood compatibility of polyethersulfone hollow fiber plasma separator].
pubmed:affiliation
Division of Hematology, First Affiliated Hospital, West China University of Medical Sciences, Chengdu 610041.
pubmed:publicationType
Journal Article, English Abstract, Research Support, Non-U.S. Gov't