pubmed:abstractText
The design, construction, and performance of a low-inductance solenoidal coil with high B(0) homogeneity for fast-field-cycling NMR is presented. It consists of six concentric layers. The conductor width is varied to minimize the B(0) inhomogeneity in the volume of the sample. This is done using an algorithm which takes the real shape of the conductor directly into account. The calculated coil geometry can be manufactured easily using standard computerized numeric control equipment, which keeps the costs low. The coil is liquid cooled and produces a B(0) field of 0.95 T at 800 A. The field inhomogeneity in a cylindrical volume (diameter 5 mm, length 10 mm) is about 10 ppm, and the inductance is 190 microH. Switching times below 200 micros can be achieved. During 6 months of operation the coil has shown good stability and reliability.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
1090-7807
pubmed:author
pubmed:copyrightInfo
Copyright 2001 Academic Press.
pubmed:issnType
Print
pubmed:volume
149
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
22-8
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Magnet design with high B(0) homogeneity for fast-field-cycling NMR applications.
pubmed:affiliation
Fachbereich Physik, TU Darmstadt, Darmstadt, 64289, Germany.
pubmed:publicationType
Journal Article