Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2-3
pubmed:dateCreated
2001-2-23
pubmed:abstractText
The performance of four commercially available ICPMS instruments of three different types was compared by means of uranium "isotopic measurements". Examined were two quadrupole sector (different generation, different manufacturer), one single detector double focusing magnetic sector and one multiple collector double focusing magnetic sector instruments. The same samples of the IRMM-072 series were used under routine conditions to measure the 233U/235U and the 233U/238U ratios which, in these samples, vary over almost three orders of magnitude from approximately 1 to approximately 2 x 10(-3). Within expanded (k = 2) uncertainties, good agreement was observed between the certified values and the data internally corrected for mass-discrimination effects. The magnitude of the evaluated uncertainties was different for each type of instrument. With the multiple collector instrument, expanded uncertainties varied from +/- 0.04% to +/- 0.24% for the 233U/235U ratio, and from +/- 0.08% to +/- 0.27% for the 233U/238U ratio. They were approximately 1 to 5 times larger with the single detector magnetic sector instrument, and approximately 10 to 25 times larger with both quadrupole sector instruments. With the multiple collector instrument, repeatability of the measurements seemed to be limited by the difficulty of correcting properly for instrumental background, whereas with the single detector magnetic sector instrument the counting statistics was the only limitation (on smallest ratios). Apparent mass-discrimination was clearly found to be larger but more reproducible (and hence easier to correct for) in the case of magnetic sector instruments than for both quadrupole sector instruments. If space charge effects were the main source of mass-discrimination for all instruments, these results are in contradiction with the hypothesis of the size of mass-discrimination decreasing with the acceleration voltage. With the single detector magnetic sector instrument in particular (when operated by changing the ion energy only), our results pointed at more than only one major source of mass-discrimination, with variable size depending on the ratios measured.
pubmed:language
eng
pubmed:journal
pubmed:status
PubMed-not-MEDLINE
pubmed:issn
0937-0633
pubmed:author
pubmed:issnType
Print
pubmed:volume
368
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
148-55
pubmed:dateRevised
2003-10-31
pubmed:articleTitle
Comparative performance study of ICP mass spectrometers by means of U "isotopic measurements".
pubmed:affiliation
Institute for Reference Materials and Measurements, Isotope Measurements Unit, European Commission--Joint Research Center, Geel, Belgium. quetel@irmm.jrc.be
pubmed:publicationType
Journal Article