pubmed:abstractText
High rates of aseptic loosening have been reported for microstructured hydroxyapatite-coated acetabular components. A macrostructured component surface (arc-deposition) not only improves resistance to shear forces experienced by the acetabular component and increases initial stability, but also provides channels for bone ingrowth. The purpose of this investigation was to radiographically compare a series of grit-blasted (microstructured) and arc-deposited (macrostructured) hydroxyapatite-coated acetabular components. A minimum 4-year retrospective radiographic analysis of acetabular components was performed on a total of 50 total hip arthroplasties. At 4 years, arc-deposited components were associated with fewer radiolucent lines in all Charnley zones, particularly Charnley zone III. While the 4-year results for arc-deposited hydroxyapatite acetabular components are superior to their microstructured predecessors, long-term results are still unknown.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0018-5647
pubmed:author
pubmed:issnType
Print
pubmed:volume
59
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
144-8
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
Radiographic comparison of grit-blasted hydroxyaptite and arc-deposited hydroxyapatite acetabular components. A four-year follow-up study.
pubmed:affiliation
Department of Orthopaedic Surgery, New York University-Hospital for Joint Diseases, New York, New York, USA.
pubmed:publicationType
Journal Article