2001-1-4
|
pubmed:abstractText |
The purpose of this study was to evaluate a large-area, flat-panel X-ray detector, which uses cesium-iodide (CsI) and amorphous silicon (a-Si). Conventional images were compared with digital images acquired with equal dose (2.5 microGy) and with 50% dose reduction. Fifteen consecutive patients were studied prospectively using an asymmetric screen-film system (detector dose, 2.5 microGy). Digital images were taken from the same patients in a posteroanterior view with detector doses of 2.5 and 1.25 microGy, respectively. The CsI/a-Si active-matrix imager had a panel-size of 43 x 43 cm, a matrix of 3 x 3k, and a pixel-pitch of 143 microm. Hard copies were presented in a random order to eight independent observers, who rated image quality according to six subjective quality criteria. Statistical significance of differences was evaluated with Student's t test for paired samples (confidence level, 95%). Digital radiographs with 2.5 and 1.25 microGy were superior to conventional images regarding all quality criteria. Statistically significant differences were observed for five of six criteria at a detector dose of 2.5 microGy and for only one quality feature at 1.25 microGy. Flat-panel digital imagers based on CsI/a-Si technique have the potential to replace conventional systems and might allow a reduction of radiation dose by 50% without loss of image quality.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Jul
|
pubmed:issn |
0883-5993
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
15
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
157-61
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:10928606-Adult,
pubmed-meshheading:10928606-Cesium,
pubmed-meshheading:10928606-Female,
pubmed-meshheading:10928606-Humans,
pubmed-meshheading:10928606-Iodine,
pubmed-meshheading:10928606-Male,
pubmed-meshheading:10928606-Middle Aged,
pubmed-meshheading:10928606-Prospective Studies,
pubmed-meshheading:10928606-Radiation Dosage,
pubmed-meshheading:10928606-Radiographic Image Enhancement,
pubmed-meshheading:10928606-Radiography, Thoracic,
pubmed-meshheading:10928606-Silicon
|
pubmed:year |
2000
|
pubmed:articleTitle |
Chest radiography with a large-area detector based on cesium-iodide/amorphous-silicon technology: image quality and dose requirement in comparison with an asymmetric screen-film system.
|
pubmed:affiliation |
Department of Diagnostic Radiology, University Hospital Regensburg, Germany.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Evaluation Studies
|