Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
2000-9-6
pubmed:abstractText
A laser-induced fluorescence detector for liquid chromatography was developed. This detector was assessed by utilizing it in conjunction with gel filtration chromatography. Using the 488 nm line of an argon ion laser for excitation and monitoring the emitted fluorescence centering at 535 nm, the limit of detection of fluorescein was 580 fM. Bovine serum albumin labeled with fluorescein was detected at a concentration of 500 fM.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0269-3879
pubmed:author
pubmed:issnType
Print
pubmed:volume
14
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
156-9
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
Laser-induced fluorescence detector for liquid chromatography: applications to protein analysis.
pubmed:affiliation
Department of Chemistry, University of Winnipeg, MB, Canada.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't