After cemented total hip arthroplasty (THA) there may be failure at either the cement-stem or the cement-bone interface. This results from the occurrence of abnormally high shear and compressive stresses within the cement and excessive relative micromovement. We therefore evaluated micromovement and stress at the cement-bone and cement-stem interfaces for a titanium and a chromium-cobalt stem. The behaviour of both implants was similar and no substantial differences were found in the size and distribution of micromovement on either interface with respect to the stiffness of the stem. Micromovement was minimal with a cement mantle 3 to 4 mm thick but then increased with greater thickness of the cement. Abnormally high micromovement occurred when the cement was thinner than 2 mm and the stem was made of titanium. The relative decrease in surface roughness augmented slipping but decreased debonding at the cement-bone interface. Shear stress at this site did not vary significantly for the different coefficients of cement-bone friction while compressive and hoop stresses within the cement increased slightly.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
0301-620X
pubmed:author
pubmed:issnType
Print
pubmed:volume
82
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
297-303
pubmed:dateRevised
2010-11-10
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
The fixation of the cemented femoral component. Effects of stem stiffness, cement thickness and roughness of the cement-bone surface.
pubmed:affiliation
Orthopaedic Hospital, Lausanne, Switzerland.
pubmed:publicationType
Journal Article, Comparative Study