The purpose of our study was to evaluate the interdependency of spatial resolution, image reconstruction artifacts, and radiation doses in virtual CT colonoscopy by comparing various CT scanning protocols. A pig's colon with several artificial polypoid lesions was imaged after air insufflation with helical CT scanning using 1-, 3-, and 5-mm collimation, and pitch values varying from 1.0 to 3.0. Virtual endoscopic images and "fly through" sequences were calculated on a Sun Sparc 20 workstation (Navigator Software, GE Medical Systems, Milwaukee, Wis.). Several reconstruction artifacts as well as overall image quality were evaluated by three independent reviewers. In addition, radiation doses for the different CT protocols were measured as multiple-scan average dose using a 10-cm ion chamber and a standard Plexiglass body phantom. Generally, image quality and reconstruction artifacts were less affected by pitch values than by beam collimation. Thus, narrow beam collimation at higher pitch values (e. g. 3 mm/2.0) seems to be a reasonable compromise between quality of virtual endoscopic images and radiation dose load.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0938-7994
pubmed:author
pubmed:issnType
Print
pubmed:volume
10
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
183-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
Virtual computed tomography colonoscopy: artifacts, image quality and radiation dose load in a cadaver study.
pubmed:affiliation
Department of Radiology, University Hospital of Innsbruck, Anichstrasse 35, A-6020 Innsbruck, Austria.
pubmed:publicationType
Journal Article, Comparative Study