To describe imaging findings of hepatic hemangiomas on power Doppler (PD) ultrasound (US) with revised Doppler parameters for preventing PD artefacts from stationary hyperechoic tissue, we prospectively evaluated 48 hemangiomas by PD US with predetermined PD settings to prevent artefactual signals from stationary hyperechoic tissue (pulse repetition frequency of 1000 Hz, medium wall filter, and PD gain of 60-85%). Intratumoral PD signals were not seen in 32 lesions (67%). Minimal (n = 15) or moderate (n = 1) intratumoral PD signals were seen in 16 lesions (33%) and were distributed in the peripheral portion only in 12 lesions (75%) and in the peripheral and central portion in 4 lesions (25%). Due to the lack of sensitivity of PD to detect slow flow in hemangiomas, PD US should no longer be used for the evaluation of echogenic liver masses caused by hemangiomas from other hypovascular malignant lesions of the liver.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0301-5629
pubmed:author
pubmed:issnType
Print
pubmed:volume
25
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1055-61
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1999
pubmed:articleTitle
Signal from hepatic hemangiomas on power Doppler US: real or artefactual?
pubmed:affiliation
Department of Radiology and Institute of Radiation Medicine, Seoul National University College of Medicine, Korea.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't