Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
8
pubmed:dateCreated
1999-10-13
pubmed:abstractText
Lithium fluoride is a commonly used solid state dosimeter. During irradiation, electrons and holes become trapped in crystal imperfections; thermoluminescence dosimetry measures their thermally induced recombination. Electron paramagnetic resonance (EPR) spectroscopy can be used to measure the resonant absorption of microwaves by the unpaired electrons trapped in LiF. In an effort to extend the use of LiF dosimeters to smaller sizes and to the harsh environments encountered in internal dosimetry, EPR was evaluated as an alternative technique to read the radiation dose delivered to TLD-100 dosimeters. TLD-100 rods were irradiated with a 60Co source to doses of 10 Gy to 100 Gy. A radiation-induced signal (with a g-value of 2.002) could be detected only at liquid nitrogen temperatures at doses above 20 Gy. The EPR spectrum of irradiated LiF contains three components, one of which correlates positively with dose. However, the low sensitivity of the technique, and difficulty in interpreting the EPR spectrum from polycrystalline dosimeters, preclude its use as a dosimetry technique.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
0031-9155
pubmed:author
pubmed:issnType
Print
pubmed:volume
44
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
2063-9
pubmed:dateRevised
2000-12-18
pubmed:meshHeading
pubmed:year
1999
pubmed:articleTitle
Feasibility of reading LiF thermoluminescent dosimeters by electron spin resonance.
pubmed:affiliation
Physics Department, London Regional Cancer Centre, Cancer Care Ontario, Canada.
pubmed:publicationType
Journal Article