pubmed-article:2801211pubmed:dateCreated1989-11-6lld:pubmed
pubmed-article:2801211pubmed:abstractTextA recently available chemiluminescence immunoassay (LIA-MAT) for TSH, set up by Byk-Sangtec Diagnostica, has been evaluated and compared with IRMA methods. The system LIA-MAT uses two monoclonal antibodies: one coated on tubes and the other labelled with isoluminol. To compute the within-assay precision profile, we estimated the response error relationship from all the duplicates (420) of eleven experiments. The CV of the response was 4-5% from the maximal response until 10,000 RLU (corresponding to about 1 microIU/ml); in the lower response range the CV worsened up to 8%. The sensitivity, derived from the precision profile, was 0.052 microIU/ml similar to that found in IRMA-MAT (0.044 microIU/ml); the working range extended from 0.33 to 100 microIU/ml. Results from LIA-MAT in the concentration range 1-30 microIU/ml were compared with the consensus mean produced by users of IRMA methods (IRMA-MAT, IRMA-Behring, Maiaclone Serono) participating in an inter-laboratory survey; a good correlation with IRMA techniques was found. The distribution of TSH determinations produced by LIA-MAT on 62 low concentration sample (less than 0.3 microIU/ml) from patients unresponsive to TRH test, was found similar to that observed for the kit IRMA Boots-Celltech assumed as reference.lld:pubmed
pubmed-article:2801211pubmed:languageenglld:pubmed
pubmed-article:2801211pubmed:journalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:2801211pubmed:citationSubsetIMlld:pubmed
pubmed-article:2801211pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:2801211pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:2801211pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:2801211pubmed:statusMEDLINElld:pubmed
pubmed-article:2801211pubmed:monthJullld:pubmed
pubmed-article:2801211pubmed:issn0884-3996lld:pubmed
pubmed-article:2801211pubmed:authorpubmed-author:PiloAAlld:pubmed
pubmed-article:2801211pubmed:authorpubmed-author:ZucchelliG...lld:pubmed
pubmed-article:2801211pubmed:authorpubmed-author:MasiniSSlld:pubmed
pubmed-article:2801211pubmed:authorpubmed-author:FerdeghiniMMlld:pubmed
pubmed-article:2801211pubmed:authorpubmed-author:CalvoSSlld:pubmed
pubmed-article:2801211pubmed:issnTypePrintlld:pubmed
pubmed-article:2801211pubmed:volume4lld:pubmed
pubmed-article:2801211pubmed:ownerNLMlld:pubmed
pubmed-article:2801211pubmed:authorsCompleteYlld:pubmed
pubmed-article:2801211pubmed:pagination185-90lld:pubmed
pubmed-article:2801211pubmed:dateRevised2007-11-15lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:meshHeadingpubmed-meshheading:2801211-...lld:pubmed
pubmed-article:2801211pubmed:year1989lld:pubmed
pubmed-article:2801211pubmed:articleTitleA new luminescence immunoassay for thyrotropin using coated tubes: evaluation and comparison with immunoradiometric assay.lld:pubmed
pubmed-article:2801211pubmed:affiliationC.N.R. Institute of Clinical Physiology, Pisa, Italy.lld:pubmed
pubmed-article:2801211pubmed:publicationTypeJournal Articlelld:pubmed
pubmed-article:2801211pubmed:publicationTypeComparative Studylld:pubmed