pubmed-article:9832377 | pubmed:abstractText | A new high-performance liquid chromatographic method for simultaneous determination of primidone (PRM) and of its three major metabolites, phenobarbital (PB), p-hydroxyphenobarbital (p-HO-PB) and phenylethylmalonamide (PEMA), in rat urine, was developed. After acid hydrolysis, these compounds were extracted from urine by means of a Bond Elut Certify LRC column with good clean-up. The extracts were chromatographed on a C18 reversed-phase column using isocratic elution at 40 degrees C, with UV detection at 227 nm. The limit of detection was 0.5 mg/ml for the four compounds. Good linearity (r2>0.99) was observed within the calibration ranges studied: 37.4-299.3 microg/ml for PRM, 26.4-211.2 microg/ml for PB, 12.5-100.2 microg/ml for p-HO-PB and 12.1-97.0 microg/ml for PEMA. Repeatability was in the range 3.1-6.8%. This method constitutes a useful tool for studies on the influence of various parameters on primidone metabolism. | lld:pubmed |