pubmed-article:4043824pubmed:abstractTextThe authors describe 4 cases of falsely recorded fetal heart rate pattern due to technical failure. Interaction of a fetal monitor and an infusion pump may lead to a 'false' fetal heart rate tracing, suggesting fetal compromise and masking the actual cardiac activity. The limits of indirect fetal monitoring are stressed.lld:pubmed
pubmed-article:4043824pubmed:languageenglld:pubmed
pubmed-article:4043824pubmed:journalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:4043824pubmed:citationSubsetIMlld:pubmed
pubmed-article:4043824pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:4043824pubmed:statusMEDLINElld:pubmed
pubmed-article:4043824pubmed:issn0378-7346lld:pubmed
pubmed-article:4043824pubmed:authorpubmed-author:SerrD MDMlld:pubmed
pubmed-article:4043824pubmed:authorpubmed-author:MashiachSSlld:pubmed
pubmed-article:4043824pubmed:authorpubmed-author:RabinoviciJJlld:pubmed
pubmed-article:4043824pubmed:authorpubmed-author:ShalevJJlld:pubmed
pubmed-article:4043824pubmed:issnTypePrintlld:pubmed
pubmed-article:4043824pubmed:volume20lld:pubmed
pubmed-article:4043824pubmed:ownerNLMlld:pubmed
pubmed-article:4043824pubmed:authorsCompleteYlld:pubmed
pubmed-article:4043824pubmed:pagination52-5lld:pubmed
pubmed-article:4043824pubmed:dateRevised2004-11-17lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:meshHeadingpubmed-meshheading:4043824-...lld:pubmed
pubmed-article:4043824pubmed:year1985lld:pubmed
pubmed-article:4043824pubmed:articleTitle'False' fetal heart rate recording caused by interaction of an infusion pump and a fetal monitor.lld:pubmed
pubmed-article:4043824pubmed:publicationTypeJournal Articlelld:pubmed
pubmed-article:4043824pubmed:publicationTypeCase Reportslld:pubmed