pubmed-article:21309477 | pubmed:abstractText | This study compared the ability of the Zeus multifunctional anaesthesia system to control haemodynamic response to surgical stimulation in semi-closed (SCA) or closed circuit anaesthesia (CCA) modes. Fifty patients undergoing gynaecological surgery were randomly assigned to SCA or CCA. Anaesthesia was induced with 2 mg propofol and 0.9 mg/kg rocuronium, intravenously, and maintained using sevoflurane (minimum alveolar concentration [MAC], 1.0) using 2 l/min oxygen plus 2 l/min nitrous oxide (SCA 4 l/min group) or 50% oxygen plus 50% nitrous oxide (CCA group). An increase in mean arterial pressure (MAP) > 20% above baseline in response to surgical stimulation provoked a stepwise increase in sevoflurane (1.3 MAC and then 1.6 MAC), followed by fentanyl 1 pg/kg intravenously (rescue drug). The time required for MAP to return to within 10% of baseline was significantly shorter in the CCA group (6.4 +/- 3.6 min) compared with the SCA 4 l/min group (10.2 +/- 6.0 min). The percentage of patients requiring fentanyl was significantly greater in the SCA 4 l/min group than in the CCA group. In conclusion, CCA controlled acute haemodynamic responses to surgical stimuli more successfully and rapidly than SCA 4 l/min, using a multifunctional anaesthesia machine. | lld:pubmed |