pubmed-article:17716847 | pubmed:abstractText | In the first two documents [Ph. Hubert, J.J. Nguyen-Huu, B. Boulanger, E. Chapuzet, P. Chiap, N. Cohen, P.A. Compagnon, W. Dewé, M. Feinberg, M. Lallier, M. Laurentie, N. Mercier, G. Muzard, C. Nivet, L. Valat, J. Pharm. Biomed. Anal. 36 (2004) 579-586; Ph. Hubert, J.J. Nguyen-Huu, B. Boulanger, E. Chapuzet, P. Chiap, N. Cohen, P.A. Compagnon, W. Dewé, M. Feinberg, M. Lallier, M. Laurentie, N. Mercier, G. Muzard, C. Nivet, L. Valat, E. Rozet, J. Pharm. Biomed. Anal., in press], a recent SFSTP Commission on the validation of analytical procedure has introduced a harmonized approach for the validation of analytical procedures. In order to complete this guide, the statistical methodology allowing to correctly conclude about the validity of a procedure is proposed in this third part of the guide. Indeed all the steps to obtain the decision tool namely the accuracy profile are described and illustrated step by step by a numerical example. This tool, based on the concept of total error (bias+standard deviation) build with a beta-expectation tolerance interval, allows to easily take the right decision and simultaneously minimizing the risk of the future use of the analytical procedure. | lld:pubmed |