pubmed:abstractText
Around 20% of patients with third generation implantable cardioverter defibrillators receive inappropriate therapy, usually triggered by atrial fibrillation. This is because the criteria used for ventricular tachycardia detection by current implantable cardioverter defibrillators are based on the analysis of a sequence of RR intervals and may be inappropriately satisfied by supraventricular tachyarrhythmias. Algorithms for ventricular tachycardia detection were challenged against the full RR interval sequences from 482 spontaneous episodes of atrial fibrillation and 260 spontaneous episodes of ventricular tachycardia to determine their ability to discriminate between the arrhythmias. The sensitivities and specificities of the algorithms were calculated over a wide range of programmable parameters. For a given window length and detection interval, the most stringent algorithms, that required all beats to be classified as "fast", were more specific than those allowing a proportion of "normal" intervals, even after adjustment for differing sensitivity. These differences were less marked for faster tachycardias. Specificity increased with the detection window length to a limit of approximately 18 beats. We conclude that ventricular tachycardia is detected with the highest specificity if all beats in an analyzed sequence are required to be "fast," even after lengthening of the tachycardia detection interval to maintain sensitivity. Further improvement in algorithm performance may require the incorporation of criteria such as tachycardia onset and stability.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
20
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
2975-83
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1997
pubmed:articleTitle
Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implications for device programming.
pubmed:affiliation
St. George's Hospital Medical School, London, United Kingdom.
pubmed:publicationType
Journal Article