pubmed:abstractText
A pyrolytic carbon coating was applied to F-75 chromium-cobalt-molybdenum alloy in an effort to reduce the release of corrosion products in vivo. After intramuscular implantation in the rat, a complex pattern of serum and urine concentration elevations of chromium, cobalt, and nickel was seen. The carbon-coated implants released more chromium and cobalt than uncoated controls, as seen by significantly elevated metal concentrations in serum and urine. Animals receiving carbon-coated implants showed a high rate of recurrent implant site inflammation. Neoplastic infiltration of 24 animals with coated implants, but not in any of the 16 animals which received either uncoated F-75 microsphere or poly(ethylene) particulate implants.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0021-9304
pubmed:author
pubmed:issnType
Print
pubmed:volume
21
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1213-30
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1987
pubmed:articleTitle
Release of corrosion products by F-75 cobalt base alloy in the rat. III: Effects of a carbon surface coating.
pubmed:affiliation
Department of Orthopaedic Surgery, University of Pennsylvania, Philadelphia 19104.