One technique for replacing the articular surface of the patella in total knee replacement is the use of a cementless porous-coated metal-backed polyethylene component. Anteroposterior, lateral, and Merchant or skyline radiographs in 10 cases of porous-coated metal-backed components that failed were evaluated for component alignment, component displacement, radiolucencies, loose beads, and integrity of the patellar component. Component failure was documented by surgery in all 10 cases. Seven cases of failure involved the patellar pegs or junction of the pegs with the metal backing and three cases involved the polyethylene portion. In the seven cases of peg-metal backing failure, three cases showed loose beads before failure and seven cases showed displacement of the metal backing at time of failure. All three cases of polyethylene failure showed narrowing or displacement of the polyethylene portion. Our experience shows that displacement of the metal backing, displacement of the polyethylene, and disruption of patellar pegs are seen in failed porous-coated metal-backed patellar components. Loose beads may be seen before peg-metal backing failure.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0361-803X
pubmed:author
pubmed:issnType
Print
pubmed:volume
155
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
555-8
pubmed:dateRevised
2008-2-15
pubmed:meshHeading
pubmed:year
1990
pubmed:articleTitle
Total knee replacement: radiologic findings in failure of porous-coated metal-backed patellar component.
pubmed:affiliation
Department of Radiology, Cleveland Clinic Foundation, OH 44195-5021.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't