Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
2
|
pubmed:dateCreated |
1998-8-12
|
pubmed:abstractText |
The aim of this study was to assess day-to-day variability of left ventricular (LV) volume and ejection fraction (EF) calculated from 3-dimensional echocardiography (3-DE) and to compare the reproducibility of the measurement with magnetic resonance imaging. Forty-six subjects were examined including 15 normal volunteers (group A) and 31 patients with LV dysfunction (group B). Precordial 3-DE acquisition was performed at 2 degrees rotational intervals and repeated 1 week later. Magnetic resonance imaging was performed at 0.5 T. End-diastolic and end-systolic LV volumes were derived using Simpson's rule by manual endocardial tracing of 8 equidistant parallel LV short-axis slices with 3-DE, whereas 9-mm slices were used with magnetic resonance imaging. The mean +/- SD of end-diastolic and end-systolic LV volumes (ml) and EF (%) from magnetic resonance imaging were 182 +/- 75, 121 +/- 76, and 39 +/- 18, whereas those from 3-DE were 182 +/- 76, 121 +/- 77, and 39 +/- 18 respectively. Day-to-day measurements of end-diastolic and end-systolic LV volumes, and EF on 3-DE were not significantly different as assessed with SEE (2.7, 1.1, and 2.4, respectively). Intra- and interobserver SEE for calculating end-diastolic and end-systolic LV volumes and EF for magnetic resonance imaging were 6.3, 4.7, and 2.1 and 13.6, 11.5, and 4.7, respectively, whereas those for 3-DE were 3.1, 4.4, and 2.2 and 6.2, 3.8, and 3.6, respectively. Day-to-day variability of LV volume and EF calculation on 3-DE were small and not significantly different for normal and dysfunctional left ventricles. Observer variabilities of 3-DE were fewer than those of magnetic resonance imaging. Therefore, 3-DE is recommended for serial assessment of LV volume and EF in normal and abnormally shaped ventricles.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jul
|
pubmed:issn |
0002-9149
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:day |
15
|
pubmed:volume |
82
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
209-14
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:9678293-Adult,
pubmed-meshheading:9678293-Aged,
pubmed-meshheading:9678293-Circadian Rhythm,
pubmed-meshheading:9678293-Echocardiography, Three-Dimensional,
pubmed-meshheading:9678293-Female,
pubmed-meshheading:9678293-Humans,
pubmed-meshheading:9678293-Magnetic Resonance Imaging,
pubmed-meshheading:9678293-Male,
pubmed-meshheading:9678293-Middle Aged,
pubmed-meshheading:9678293-Reproducibility of Results,
pubmed-meshheading:9678293-Stroke Volume,
pubmed-meshheading:9678293-Ventricular Dysfunction, Left
|
pubmed:year |
1998
|
pubmed:articleTitle |
Measurements and day-to-day variabilities of left ventricular volumes and ejection fraction by three-dimensional echocardiography and comparison with magnetic resonance imaging.
|
pubmed:affiliation |
Heart Center and Thoraxcenter, Division of Cardiology, University Hospital Rotterdam-Dijkzigt, Erasmus University, Rotterdam, The Netherlands.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't
|