pubmed:abstractText |
Interference by salicylic acid was noted in a high pressure liquid chromatographic assay of theophylline in serum. The acid eluted very close to theophylline in a mobile phase consisting of 0.01 M acetate buffer, pH 4.0, containing 28% methanol on a C-18 reverse-phase column. The two peaks could be resolved by switching to a mobile phase containing 18% methanol, 1.6% acetonitrile, and 1.6% acetic acid in water. However, in this mobile phase traces of the extracting solvent, ethyl acetate, caused a sharpening of the theophylline peak leading to spuriously high results. The problem was overcome by using chloroform to extract the theophylline from serum.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical |
http://linkedlifedata.com/resource/pubmed/chemical/Acetates,
http://linkedlifedata.com/resource/pubmed/chemical/Salicylic Acid,
http://linkedlifedata.com/resource/pubmed/chemical/Salicylic Acids,
http://linkedlifedata.com/resource/pubmed/chemical/Theophylline,
http://linkedlifedata.com/resource/pubmed/chemical/ethyl acetate
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0163-4356
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
4
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
323-4
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading | |
pubmed:year |
1982
|
pubmed:articleTitle |
Interferences in a high pressure liquid chromatographic assay of theophylline.
|