Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
1983-10-8
pubmed:abstractText
The transfer characteristic of a 16 channel ink-jet writer used in a cardiovascular laboratory is described. The test method for measuring linearity, gain and phase response is described and results given. The transfer function was calculated from the normalized experimental results and allows prediction of the behaviour of the writer for different input signals. It is shown that the ink-jet galvanometers are linear for signal input amplitudes between 0 and 2.5 V, and have a flat frequency response between DC and 640 Hz (+/- 1%).
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0141-5425
pubmed:author
pubmed:issnType
Print
pubmed:volume
5
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
257-61
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1983
pubmed:articleTitle
Transfer characteristics of an ink-jet writer for cardiovascular studies.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't