pubmed:abstractText |
This research was initiated in order to study the extraction forces associated with solid, single-vented and multi-vented blade dental implants placed in mandibles of dog for 2 months. An extraction device was used to measure the extraction forces on five of the implants and a histological study was carried out around the sixth multi-vented implant. The main finding was that the extraction forces for the multi-vented blade implants were 6 to 12 times those of the single-vented and solid blades. Other findings in this study were: none of the implants exhibited mobility; sulci ranged from 2 to 3 mm at the implant neck; bone covered the implant shoulder. There was gross and histologic evidence of bone growth in the vents of the multi-vented blade; close adaptation of the bone, connective tissue capsule and implant; and no gross, histologic, clinical or radiographic evidence of bone pathology.
|