1985-3-21
|
pubmed:abstractText |
Thresholds in constant current and constant voltage are reported for 405 CPI 4118 (Cardiac Pacemakers, Inc.) tined leads. In 98% of the cases, the constant current threshold was equal to or lower than 0.6 mA. A constant voltage threshold was equal to or lower than 0.5 V in 96% of the cases. Three external pacemakers and two pacing system analyzers were evaluated as threshold testing devices. For analyzing the waveforms of current and voltage stimuli, a calibrated isolation amplifier was used. We could not find a definite difference between the value, at which pacing is lost with reduction of the pulse generator output, and the value at which increase reestablishes pacing.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0147-8389
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
8
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
66-72
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
1985
|
pubmed:articleTitle |
Measurement of pacing threshold.
|