Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
2008-12-26
pubmed:abstractText
A polytetrafluoroethylene (PTFE)-covered stent is specially used to treat coronary perforation complicating percutaneous intervention in order to prevent the aneurysm from rupturing, but until now it has not been known if endothelialization occurs inside this type of stent. A patient with a giant aneurysm of the right coronary artery underwent successful implantation of a PTFE-covered stent. Angiography at 9-month follow-up showed focal restenosis at the proximal edge of the stent and coronary angioscopy revealed restenosis as a result of thrombus formation. Absence of endothelialization in the covered stent was also detected by angioscopy and optical coherence tomography. These findings suggest that in-stent thrombosis must be prevented after PTFE-covered stent implantation.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
1346-9843
pubmed:author
pubmed:issnType
Print
pubmed:volume
73
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
190-3
pubmed:meshHeading
pubmed:year
2009
pubmed:articleTitle
Delayed endothelialization after polytetrafluoroethylene-covered stent implantation for coronary aneurysm.
pubmed:affiliation
Division of Cardiology, Nippon Medical School, Tokyo, Japan. takanom@nms.ac.jp
pubmed:publicationType
Journal Article, Case Reports