pubmed:abstractText
Endotracheal tube leaks (ETTLs) occur in neonates ventilated with uncuffed tubes. Assuming that the influence of ETTLs might be neglected during expiration, only expiratory tidal volume is measured for calculation of expiratory compliance in cases of large ETTLs. However, expiratory ETTL might be substantial. Therefore, we evaluated the effect of ETTL size on expiratory tidal volume and compliance.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
1529-7535
pubmed:author
pubmed:issnType
Print
pubmed:volume
9
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
69-75
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Limitation of measurements of expiratory tidal volume and expiratory compliance under conditions of endotracheal tube leaks.