pubmed:abstractText
Parallel, spatial-encoded MRI requires a large number of independent detectors that simultaneously acquire signals. The loop structure and mutual coupling in conventional phased arrays limit the number of coils and therefore the potential reduction in minimum scan time achievable by parallel MRI tchniques. A new near-field MRI detector array, the planar strip array (PSA), is presented that eliminates the coupling problems and can be extended to a very large number of detectors and high MRI frequencies. Its basic structure is an array of parallel microstrips with a high permittivity substrate and overlay. The electromagnetic (EM) wavelength can be adjusted with the permittivity, and the strip lengths tuned to a preselected fraction of the wavelength of the MRI frequency. EM wave analysis and measurements on a prototype four-element PSA reveal that the coupling between the strips vanishes when the strip length is either an integer times a quarter wavelength for a standing-wave PSA, or a half wavelength for a travelling-wave PSA, independent of the spacing between the strips. The analysis, as well as phantom and human MRI experiments performed by conventional and parallel-encoded MRI with the PSA at 1.5 T, show that the decoupled strips produce a relatively high-quality factor and signal-to-noise ratio, provided that the strips are properly terminated, tuned, and matched or coupled to the preamplifiers. Magn Reson Med 45:673-683, 2001.
pubmed:grant
pubmed:commentsCorrections
http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-10542355, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-10680691, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-10800037, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-1745117, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-2046612, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-2067400, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-2266841, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-7968443, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-8978639, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-9324327, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-9702709, http://linkedlifedata.com/resource/pubmed/commentcorrection/11283996-9877459
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0740-3194
pubmed:author
pubmed:copyrightInfo
Copyright 2001 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
45
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
673-83
pubmed:dateRevised
2011-5-3
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Planar strip array (PSA) for MRI.
pubmed:affiliation
Division of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA. lee@crd.ge.com
pubmed:publicationType
Journal Article, Comparative Study, Research Support, U.S. Gov't, P.H.S.