Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
2001-1-31
pubmed:abstractText
Near infrared spectroscopy is an increasingly important tool for the investigation of human brain function, however, to date there have been few systematic evaluations of accompanying thermal effects due to absorption. We have measured the spatial distribution of temperature changes during near infrared irradiation (789 nm) as a function of laser power, in both excised tissue (chicken meat and skin) and in the forearm of an awake human volunteer. Light was applied using a 1 mm optical fiber which is characteristic of the topographic system. The temperature of excised chicken tissue increased linearly with power level as 0.097 and 0.042 degrees C/mW at depths of 0 and 1 mm, respectively. Human forearm studies yielded temperature changes of 0.101, 0.038, and 0.030 degrees C/mW at depths of 0.5, 1.0, and 1.5 mm, respectively. Due to direct irradiation of the thermocouple all measurements represent the maximum temperature increase from the laser. In all cases the estimated heating effects from continuous wave optical topography systems were small and well below levels which would endanger tissue cells. The close similarity between ex vivo and in vivo measurements suggests negligible contributions from blood flow in the skin which was further supported by measurements during cuff ischemia. Heating effects decreased sharply with both depth and lateral position; thus, for optode spacings greater than a few millimeters, fibers can be treated independently. Finite element analysis confirms that the experimental results are consistent with a simple heat conduction model.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
1083-3668
pubmed:author
pubmed:issnType
Print
pubmed:volume
5
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
383-90
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
Assessment of heating effects in skin during continuous wave near infrared spectroscopy.
pubmed:affiliation
Hitachi Central Research Laboratory, Kokubunji-shi, Tokyo, Japan.
pubmed:publicationType
Journal Article, Comparative Study, Evaluation Studies