An absorbed dose in the narrow beam is calculated based on the depth dose distribution and a field factor. The field factor has to be measured with especially high accuracy because it is highly variable. The field factor was calculated when detection size was changed, by using Monte Carlo simulation, which had no energy dependency or geometrical error. Then the relation between field factor and detection size in the narrow beam was investigated. An absorbed dose in peak depth and reference depth according to detection size was calculated for each field size. Detection size dependency was recognized in the case of measuring a field factor, because the absorbed dose tended to decrease as detection size increased. The absorbed dose in the narrow beam has to be calculated within a change of +/- 2%. The change in peak depth according to detection size was eliminated, and then the relation between an absorbed dose at the ideal detection size of 0 mm phi by extrapolation and detection size which has a deference of 2% from it, were acquired. In addition, the maximum usable detection size was estimated in the case of measuring the field factor.
pubmed:language
jpn
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0048-0428
pubmed:author
pubmed:issnType
Print
pubmed:volume
59
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
338-41
pubmed:dateRevised
2011-7-26
pubmed:meshHeading
pubmed:year
1999
pubmed:articleTitle
[Detection size dependence of field factor for narrow beam].
pubmed:affiliation
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences.